Weather: high humidity, hot
- Fill the li7700 water bucket
It was 10% left
Done
Done, we empty USB
- Download met data and remove program for eclipse
Done
- Change position and config of phenocam
Done
CTD sensor is above water table so showing no data (conductivity and salinity reading 0), we measured the water level is 3cm high from the bottom of CTD.
- The fan inside HMP stop working again
- We cannot connect to the datalogger through ethernet, we connect from the router, but the download is super slow. If we cannot connect manually, check wireguard and deactive it.
- Next time bring a spouge to clean the sticked bird pop
- No standing water in the field, it looks dry (yellow) compare to field nearby and full of weeds.
Vegetation height: 40, 90, 70, 52, 75, 60
Readings:
| |
Before cleaning sensors |
After cleaning sensors |
| LI-7500 readings |
| CO2 ppm |
393.29 |
391.18 |
| H20 ppm |
27.54 |
28.09 |
| T ºC |
29.27 |
29.80 |
| P kPa |
101.35 |
101.31 |
| SS % |
95 |
100.1 |
| LI-7700 readings |
| CH4 ppm |
1.989 |
2.001 |
| RSSI % |
23.25 |
79.3 |