Weather: sunny with breeze
A little left 5%
done
done
done
done
- Check the diagnostic table: is it stillknown (unable to contact status server)?
No, everything is ok
Datalogger is ahead, synch with pc now.
We found thermal couple TC_1_32 has higher value (35 degree) than the other two (27 degree).
We found Licor7700 has several problems, the desiccant need to be replaced. and the thermal couple inside was broken (showed 9999), this is why it showed unreasonable value (67ppm). We found everything else for Licor7700 look fine. Next time we need to fix these two. But this will not affect the data because we also have CH4 density data in m3, just when to process the data, use other temperature to convert it to ratio.
Vegetation height:
Readings:
| |
Before cleaning sensors |
After cleaning sensors |
| LI-7500 readings |
| CO2 ppm |
397 |
409.86 |
| H20 ppm |
519.7 |
619.2 |
| T ºC |
21.07 |
20.43 |
| P kPa |
101.37 |
101.36 |
| SS % |
85.35 |
98.034 |
| LI-7700 readings |
| CH4 ppm |
65.36 |
67.1067 |
| RSSI % |
9.97 |
81.15 |