Weather: Sunny day with some fluffy clouds in the sky. Sea breeze but it feels very hot.
The site remains green but the phragmites have already some yellow leaves. Pep and Judith are in the lagoon collecting data from the water sensors. A car from Seo birdlife with Rietvell volunteers was at the site. They must be doing bird census.
Wash reservoir: it had the metallic lid off, probably pep forgot to close it. Regardless, it was 25% full. I refilled it.
Met data: Angèlica removed the precipitation bucket last time. Now we don’t have a sensor, but it keeps reading 0, so it might be the wiring. The rest of variables look good.
Have checked table status, let’s keep the tables as are, then at the start of the new year we can reset them, see if memory holds.
Flux data: problems connecting to the smartflux unit. Lights blink though, showing it is running. Tried 3 times, didn’t work, so I restarted the smartflux unit. After restarting it worked.
When I accessed the unit, there was a warning on the biomet system. Problems contacting it, but ip addresses is ok and data seems ok too.
Downloaded raw, results and summaries from august. No data for September. Data stops 8th of august.
There were some spiderwebs in the box and on the sensors but nothing outrageous. Checked sensor connections are tight and zip-tided solar panel cables. At 9.52 went to help out Qin finish the sensors.
Readings:
| |
Before cleaning sensors |
After cleaning sensors |
| LI-7500 readings |
| CO2 ppm |
406.46 |
395.49 |
| H20 ppm |
967.4 |
1045 |
| T ºC |
26.64 |
26.9 |
| P kPa |
101.52 |
101.59 |
| SS % |
80.16 |
98.9 |
| LI-7700 readings |
| CH4 ppm |
1.98 |
2.01 |
| RSSI % |
11.3 |
81.5 |